Gas Cooler Series TC-Standard+

Gas Cooler Series TC-Standard+

  • Low washout effects
  • Compact design: Pre-installed and ready to connect
  • Low maintenance costs based on easy accessibility
  • Optimised heat exchanger type 2 in Duran glass or PVDF
  • Adjustable outlet dew point and alarm thresholds
  • Nominal cooling capacity 100 kJ/h (40 °C version) or 90 kJ/h (50 °C version)
  • Dew point stability 0.1 °C
  • Status display and output
  • Cooling block temperature display
  • Optional 4 - 20 mA or Modbus RTU signal output
  • Optional CE mark or FM approval
  • Moisture detector, filter and condensate pump optional

Many analysis processes require extracting sample gas from the process. This also extracts process-related contamination such as particles or moisture. These can impact the measurement results or damage the measuring cells. The sample gas must therefore be conditioned before entering the analyser. The sample gas cooler reduces the gas temperature to below the dew point for this purpose, causing moisture to drop out, which is then discharged as condensate.

In addition to the status output to monitor the sample gas cooler function, we offer an optional 4 – 20 mA analog output or digital interface. The process control can access the process and diagnostic data via the Modbus RTU interface as well as configure the device settings.

The TC-Standard+ series features a new generation heat exchangers with a particularly low wash out effect of water- soluble components and are specifically suitable for measuring emissions. Particularly the wash out effect of SO2 is low. These coolers can therefore be used for so-called automated measuring systems (AMS) per EN 15267-3.